MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7005
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in left anterior descending artery.A 3.0mm x 8mm quantum maverick balloon catheter was advanced for use.However, during the procedure, the delivery shaft was fractured.The device was completely removed normally from the patient's body.The procedure was completed with another of the same device.There were no complications reported and the patient was stable.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in left anterior descending artery.A 3.0mm x 8mm quantum maverick balloon catheter was advanced for use.However, during the procedure, the delivery shaft was fractured.The device was completely removed normally from the patient's body.The procedure was completed with another of the same device.There were no complications reported and the patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter and photo media.The device was visually and microscopically examined.There was a separation of the hypotube at 78.8cm from the strain relief.The separated ends of the hypotube were ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and the balloon.There was blood in the guidewire lumen and the balloon folds of the tightly folded balloon.Inspection of the remainder of the device presented no other damage or irregularities.Photo media was reviewed and found to be aligned to what was seen in product analysis as the separation was found in both analysis, and media review.Product analysis confirmed the reported break as there was separation to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14597481
Report Number2134265-2022-05760
Device Sequence Number2098352
Product Code LOX
UDI-Device Identifier08714729392569
UDI-Public08714729392569
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/18/2023
Device Model Number7005
Device Catalogue Number7005
Device Lot Number0026381992
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexFemale
Patient Weight78 KG
-
-