• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYLIS MEDICAL COMPANY INC. VERSACROSS TRANSSEPTAL SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYLIS MEDICAL COMPANY INC. VERSACROSS TRANSSEPTAL SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number VXS85-35-63-45-D0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pericardial Effusion (3271)
Date of Event 05/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A case of pericardial effusion requiring intervention was reported during a left atrial appendage closure (laac) procedure, where the baylis medical, versacross steerable sheath (vst), versacross transseptal sheath (vxs) and versacross rf wire (vxw) were used for the initial transseptal puncture.The vst was wired with the vxw and inserted in the patient.Due to complex patient anatomy, the physician experienced difficulty maneuvering the transesophageal echocardiography (tee) probe to visualize tenting of the septum.The physician attempted to rewire and re-position the vst on the septum six more times.Tenting of the septum could not be confirmed due to poor imaging and the vst was exchanged with the vxs.The vxs was shaped, wired and inserted for positioning.With no clear imaging of the septum or the vxs, the physician requested another sized sheath.Before the sheath could be exchanged, the patient's blood pressure began to drop.A perforation leading to pericardial effusion was noted in the right atrial appendage.The laac procedure was aborted.Pericardiocentesis was performed.Cardiothoracic surgery was consulted.More fluid was removed and the patient was stabilized.There is no reasonable evidence to suggest that the baylis medical devices caused or contributed to the reported incident.However, as baylis medical devices were reported to be among the devices used in the procedure, baylis medical has decided to submit this report.A sperate mdr has been submitted for the versacross rf wire and versacross steerable sheath under mfr report numbers: 3019751610-2022-00002 and 3019751610-2022-00004 respectively.
 
Additional Manufacturer Narrative
There is no reasonable evidence to suggest that the baylis medical devices caused or contributed to the reported incident.However, as baylis medical devices were reported to be among the devices used in the procedure, baylis medical has decided to submit this report.A dhr review has been completed for the lot in question and all devices in the lot met relevant requirements prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS TRANSSEPTAL SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal, quebec H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
2775 matheson blvd. east
mississauga, ontario L4W 4 P7
CA   L4W 4P7
Manufacturer Contact
meghal khakhar
2775 matheson blvd. east
mississauga, ontario L4W 4-P7
CA   L4W 4P7
MDR Report Key14597530
Report Number3019751610-2022-00003
Device Sequence Number20213083
Product Code DYB
UDI-Device Identifier00685447003380
UDI-Public(01)00685447003380(17)230812(10)VXFA200820
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K183655
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/12/2023
Device Model NumberVXS85-35-63-45-D0
Device Lot NumberVXFA200820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2022
Initial Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-