Model Number 29462
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the outer layer was detached.A 180cm/135cm renegade hi-flo fathom system was selected for use.Prior the procedure, the device was unpacked according to the standard process, however the outer layer was detached.The procedure was completed with another of same device.No complications were reported and patient was stable post procedure.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The device was inspected for any damage or irregularities.The renegade was received with an unknown guide wire still inserted.The device was hydrated, and the guide wire was removed.One fracture was noted 53.6 cm from the hub.No other damage was noted on the device.
 
Event or Problem Description
It was reported that the outer layer was detached.A 180cm/135cm renegade hi-flo fathom system was selected for use.Prior the procedure, the device was unpacked according to the standard process, however the outer layer was detached.The procedure was completed with another of same device.No complications were reported and patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE HI-FLO FATHOM SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14598311
Report Number2134265-2022-05945
Device Sequence Number5811187
Product Code KRA
UDI-Device Identifier08714729790853
UDI-Public08714729790853
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140329
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2024
Device Model Number29462
Device Catalogue Number29462
Device Lot Number0027816929
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer07/13/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight80 KG
-
-