Model Number 3341
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 04/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Related manufacturer reference numbers: 1627487-2022-03142.It was reported the patient's scs system was autoreducing and patient was receiving inadequate therapy.Lead diagnostics indicated high impedances on multiple contacts in the cervical and lumbar leads (related manufacturer reference numbers: 1627487-2022-02396; 1627487-2022-02397).As a result, surgical intervention occurred wherein it was discovered there was a fracture of the lead extension (related manufacturer reference number: 1627487-2022-03142).Both thoracic leads and extensions were explanted.A new octrode was placed at t8 to address the issue.During the procedure, the physician replaced the ipg electively.The set screw on port 1-8 of the new ipg shifted and was repositioned and physician began closing.As the physician was completing the final sutures, the patient went into cardiac arrest and was flipped (related manufacturer reference number: 3006705815-2022-14759).Once the patient was stable, the physician completed the case with the patient laying on their side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Common Device Name
SCS EXTENSION
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14598402
Report Number1627487-2022-03141
Device Sequence Number12135525
Product Code LGW
UDI-Device Identifier05414734402279
UDI-Public05414734402279
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2016
Device Model Number3341
Device Catalogue Number3341
Device Lot Number4562614
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Initial Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR X3; SCS IPG
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight91 KG
-
-