HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL Back to Search Results
Model Number P1190A000042
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2022
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the bed had no audible brake not set alarm.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Manufacturer Narrative
The hillrom technician found the brake alarm was not working and replaced the high/low foot column which includes the brake switch.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on mar 30, 2022.It is unknown if the facility performed any other preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the high/low foot column which includes the brake switch to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, MANUAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon
Manufacturer Contact
melrita gilliam
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key14598935
MDR Text Key293452296
Report Number3006697241-2022-00035
Device Sequence Number1
Product Code FNJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1190A000042
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/27/2022
Initial Date FDA Received06/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-