• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number PLC141000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 05/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient weight: asked but unavailable.Other relevant history, including preexisting medical conditions: asked but unavailable.Concomitant medical products and therapy dates: asked but unavailable.
 
Event or Problem Description
On (b)(6) 2021 the patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.On (b)(6) 2022 the field sales associate was informed that the patient's right external iliac artery had dilated.Reportedly there was no aneurysm present in the patient's right iliac artery, but the native vessel had dilated and expanded.The physician stated that the issue was not suspected to be device related, but it was suspected that a type ii endoleak may be present.Presence of an endoleak was not confirmed.A reintervention was performed on (b)(6) 2022 to extend the contralateral leg component in the patient's right external iliac artery.Reportedly an additional plc121000 contralateral leg component was successfully implanted.There were no further reported issues.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
jenna lopez
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14599013
Report Number3013164176-2022-01397
Device Sequence Number6085590
Product Code MIH
UDI-Device Identifier00733132618637
UDI-Public00733132618637
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/20/2023
Device Model NumberPLC141000
Device Catalogue NumberPLC141000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Initial Report FDA Received Date06/03/2022
Date Device Manufactured03/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient SexMale
-
-