MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.5MM X 25MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.5MM X 25MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS45025
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 05/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device remained inside patient.
 
Event or Problem Description
It was reported that 9 months post flow diverter stent placement procedure for left ica (internal communicating artery) aneurysm, the subject device (stent) was retracted and was no longer in the position.The subject device stent moved distal and proximal tines were not as opened resulting slowed flow through the subject device and to distal vessels.The balloon angioplasty with sceptor and tacked down with atlas stent as medical intervention.The patient was discharged and doing well.
 
Event or Problem Description
It was reported that 9 months post flow diverter stent placement procedure for left ica (internal communicating artery) aneurysm, the subject device (stent) was retracted and was no longer in the position.The subject device stent moved distal and proximal tines were not as opened resulting slowed flow through the subject device and to distal vessels.The balloon angioplasty with sceptor and tacked down with atlas stent as medical intervention.The patient was discharged and doing well.
 
Additional Manufacturer Narrative
Although the dhr (device history record) could not be reviewed because the lot number remains unknown, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.5MM X 25MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14599323
Report Number3008881809-2022-00268
Device Sequence Number5901047
Product Code OUT
UDI-Device Identifier07613327375268
UDI-Public07613327375268
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFDS45025
Device Catalogue NumberFDS45025
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date06/04/2022
Supplement Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCEPTER BALLOON CATHETER
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-