• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 2MM X 6CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 2MM X 6CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036122060
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during procedure, the resistance encountered when advancing the coil (subject device) in the introducer sheath, in the hub of the microcatheter and in the microcatheter.The coil (subject device) was prematurely detached within the microcatheter.The detached coil (subject device) could be removed by retrieving the entire microcatheter.No fragments remained in the patient.The procedure was completed successfully by replacing the coil (subject device) and the microcatheter.There were no clinical consequences reported due to the event.
 
Additional Manufacturer Narrative
H4 manufacturing date; added.D4 expiration date: added.Due to the automated mes (manufacturing execution system) system, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that the device was confirmed to be in good condition during preparation/prior to use on the patient, the patient anatomy was very meandering.It was also reported that continuous flush was not set up and maintained throughout the clinical procedure.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to user error.Therefore, an assignable cause of user error will be assigned to this complaint.
 
Event or Problem Description
It was reported that during procedure, the resistance encountered when advancing the coil (subject device) in the introducer sheath, in the hub of the microcatheter and in the microcatheter.The coil (subject device) was prematurely detached within the microcatheter.The detached coil (subject device) could be removed by retrieving the entire microcatheter.No fragments remained in the patient.The procedure was completed successfully by replacing the coil (subject device) and the microcatheter.There were no clinical consequences reported due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 2MM X 6CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14599391
Report Number3008881809-2022-00269
Device Sequence Number17663725
Product Code HCG
UDI-Device Identifier07613327128260
UDI-Public07613327128260
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/24/2025
Device Model NumberM0036122060
Device Catalogue NumberM0036122060
Device Lot Number23431412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date06/04/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNKNOWN MANUFACTURER); TARGET COIL (STRYKER)
Patient SexUnknown
-
-