|
H3, h6: it was reported that a bhr revision surgery was performed for unknown reasons.No further information has been supplied and the devices, used in treatment, were not returned for analysis.As no device batch numbers were provided for investigation for the femoral head, a manufacturing record review, device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the femoral head.A review of historical complaints data was performed using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.Other similar complaints have been identified for the part number and the reported failure mode; this will continue to be monitored.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.Smith and nephew has not received the explanted device and/or adequate materials to fully evaluate the complaint from a clinical perspective.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Additionally, specific factors known to contribute to the alleged fault cannot be provided due to the insufficient information.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
|