• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH+NEPHEW LYON SCREWDRIVER TORX T7

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH+NEPHEW LYON SCREWDRIVER TORX T7 Back to Search Results
Model Number 219127ND
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during sterilization process after surgery, it was noticed that a part of the tip of the screwdriver torx t7 had broke.Procedure been completed, without any delay, with the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Correction in section e.
 
Additional Manufacturer Narrative
H3, h6: the product has been received at the lyon site for evaluation and so the complaint is confirmed.A visual inspection at receipt confirmed the breakage of the hexalobular tip of the screwdriver.It is observed that the tip is twisted at the location of the breakage.It means that a plastic deformation of the tip occurred prior to the breakage and is due to the application of a stress superior to the mechanical resistance of the device.No other anomaly is observed.The following investigative actions were performed.Refer to the respective tasks for further detail.Complaint history review: the complaint history review identify one previous similar reported event for this device and the first for the same product lot.No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.Risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.Dhr/batch record review: identified no issues or manufacturing abnormalities which could have caused or contributed to the reported event.This review determined that the device met manufacturing specifications upon release for distribution.Capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.Medical investigation monitoring board (mimb) review: per mimb, a medical investigation was not required.The complaint alleges no injury to the patient.According to risk management documentation for the hallulock instruments, application of excessive mechanical forces, improper check prior to the use or improper raw material are identified as failure mode which could result in breakage of the hallu lock driver.Based on this investigation, the need for corrective action is not indicated as no non-conformances or product deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.This investigation can be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREWDRIVER TORX T7
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SMITH+NEPHEW LYON
immeuble sequoia 2,
97 alle alexandre borodine
saint priest rhone 69800
FR  69800
Manufacturer (Section G)
SMITH+NEPHEW LYON
immeuble sequoia 2
97 alle alexandre borodine
saint priest rhone 69800
FR   69800
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14599415
Report Number3002952164-2022-00004
Device Sequence Number14927801
Product Code HXX
UDI-Device Identifier10381780052227
UDI-Public10381780052227
Combination Product (Y/N)N
Initial Reporter CountryNL
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number219127ND
Device Catalogue Number219127ND
Device Lot NumberFTL6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer05/16/2022
01/10/2023
Initial Report FDA Received Date06/04/2022
Supplement Report FDA Received Date07/26/2022
01/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-