This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: according to the information received, it was reported that ¿ it became stuck mid way through deployment and the suture would no longer pull through¿.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (8l58181), and no non-conformance was identified.As part of depuy mitek's quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
It was reported by the sales rep in australia that during an anterior cruciate ligament reconstruction procedure on an unknown date, it was observed that the speedtrap construct green/white, long suture device was stuck midway through deployment that it would no longer pull through.The suture was cut off the tendon graft; and another like device was used to complete the procedure with a delay of two minutes.There were fragments generated but were easily removed without additional intervention.There were no adverse patient consequences reported.No additional information was provided.
|