Model Number 280023MM
Medical Device Problem Code Gradient Increase (1270)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with a 23mm 2800 pericardial aortic valve underwent a valve-in-valve procedure after an implant duration of five (5) years, four (4) months due to aortic stenosis.The tavr procedure was performed with a 23mm 9750tfx transcatheter valve.The patient was noted to be stable post procedure.
 
Additional Manufacturer Narrative
Additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine 92614
Manufacturer Contact
reginald santos
1 edwards way
irvine, CA 92614
9492502731
MDR Report Key14599642
Report Number2015691-2022-06053
Device Sequence Number6113009
Product Code DYE
UDI-Device Identifier00690103021660
UDI-Public(01)00690103021660(17)200830
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date08/30/2020
Device Model Number280023MM
Device Catalogue Number280023MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/05/2022
Supplement Report FDA Received Date07/29/2022
Date Device Manufactured08/31/2016
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age62 YR
Patient SexMale
-
-