| Model Number |
H7493942832450 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that a stent dislodged.A percutaneous coronary intervention (pci) was performed on a tortuous circumflex (cx).A guidezilla, a 3.00mm x 40mm apex balloon, and a 4.50 x 32mm synergy megatron stent were selected for treatment.However, the stent came off the balloon, and the balloon was noted to be sheared.The devices were removed from the patient.The procedure was successfully completed with another of the same device.No patient complications resulted in relation to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 initial reported address 1: (b)(6).Media review: images were received from the customer, which reveal a shaft break.
|
| |
|
Event or Problem Description
|
|
It was reported that a stent dislodged.A percutaneous coronary intervention (pci) was performed on a tortuous circumflex (cx).A guidezilla, a 3.00mm x 40mm apex balloon, and a 4.50 x 32mm synergy megatron stent were selected for treatment.However, the stent came off the balloon, and the balloon was noted to be sheared.The devices were removed from the patient.The procedure was successfully completed with another of the same device.No patient complications resulted in relation to this event.
|
| |
|