MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT Back to Search Results
Model Number VENUL14080
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.Manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.The sample was not returned for evaluation and image was not provided.Therefore, the investigation is inconclusive for the reported issue.A definitive root cause could not be determined based upon the available information.A review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.The information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a stent placement, the device allegedly failed to deploy.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Common Device Name
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14600028
Report Number9681442-2022-00172
Device Sequence Number20103007
Product Code QAN
UDI-Device Identifier00801741103506
UDI-Public(01)00801741103506
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P180037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVENUL14080
Device Catalogue NumberVENUL14080
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Initial Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1897-2021
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-