| Catalog Number |
03003523001 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer determined the issue was caused by a defective chloride electrode.The analyzer was checked.The electrode was replaced by the customer and the issue was resolved.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with the chloride electrode on a cobas integra 400 plus (serial number (b)(4)).The sample initially resulted in a chloride value of 142 mmol/l and this value was reported outside of the laboratory.The sample was repeated, resulting in a value of 115 mmol/l.The repeat value was deemed correct and a corrected report was issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon customer inspection of the affected electrode, the customer stated they saw something inside the electrode that looked like fungus.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|