|
WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.8 VA LCKNG BUTTRESS PIN T8 SD REC/18; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
02.210.088 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, during an orthopedic procedure, the screws would not lock in 2 variable angle (va) distal radius plates.Multiple locking buttress pins and locking screws were tried on 2 different plates.They could not get the locking screws to engage the plate.A screw in drill guide would lock in correctly however, eventually a 6 hole lcp volar column plate was used that worked fine.Additional x-rays were taken.There was a surgical delay of more than 30 minutes.The procedure completed successfully.There was no patient consequence reported.This is report 9 of 11 for (b)(4).This report is for a buttress pin.
|
| |
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9.H3, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the 1.8 va lckng buttress pin t8 sd rec/18 was found the thread of the screw dull.This can be related with the unable to assemble complaint.No other defect was found.A dimensional inspection for the 1.8 va lckng buttress pin t8 sd rec/18 was unable to be performed due to post manufactured damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 1.8 va lckng buttress pin t8 sd rec/18 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|