MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON LEGACY 5.5; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON LEGACY 5.5; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 7480113
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding an event happened during intra-op of the reported product.The pre-op diagnosis of patient was lumbar disc herniation with radiculopathy, lumbar spinal stenosis, lumbar osteoarthropathy.It was reported that, the patient underwent lumbar internal fixation surgery in other hospital in 2012, using other brand of internal fixation product at that time.The patient underwent revision surgery on (b)(6) 2022.Internal fixation was performed with medtronic spine legacy 5.5 system screws during the operation.During the operation, the tip of the nailing device was broken, and the broken part of the nailing device remained in the inner core of the screw, so that the screw could not be fixed with the rod.Immediately, the screw was removed, and new medtronic screw was replaced for fixation.The operation was completed without adverse consequences.The procedure or technique performed was removal of lumbar internal fixation device, lumbar laminectomy and decompression, nucleus pulposus exploration, internal fixation via pedicle system, intertransverse bone graft fusion.There was no patient symptoms or complications reported.There was no additional treatment or surgery performed as a result.There was a delay of less than 60 minutes reported as a result of this event.No further complications reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON LEGACY 5.5
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key14602263
Report Number1030489-2022-00526
Device Sequence Number14743621
Product Code LXH
UDI-Device Identifier00721902861468
UDI-Public00721902861468
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7480113
Device Catalogue Number7480113
Device Lot NumberPR15M005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Initial Report FDA Received Date06/06/2022
Date Device Manufactured03/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
-
-