DEPUY INTERNATIONAL LTD - 8010379 CORAIL2 LAT COXA VARA SIZE 11; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 3L93711
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Fall (1848); Foreign Body Reaction (1868); Pain (1994); Ambulation or Postural Difficulties (2544); Inadequate Osseointegration (2646)
Date of Event 04/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
In (b)(6) 2021, during a false movement, the patient felt a sharp pain, then some sensations of cracking and grinding, with the need to walk with a cane.On the imagery examination, a fracture of the head the prosthesis.Ct scan: metallosis and early loosening of the femoral metaphysis and a pericotyloid granuloma.Explantation of the total hip prosthesis and replacement with a ceramic total hip prosthesis with pe cement.Metallosis observed.Doi: (b)(6) 2004, dor: (b)(6) 2022, right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was received for analysis.Visual examination of the device exhibited a failure of bone ingrowth onto stem.This, considering extended implanted time of device on patient's body.Based on observations of the involved devices, it was found that femoral head got fractured into several pieces leading to stem's taper articulating and applying pressure to liner surface.The taper presented a remarkable worn condition and a significant deformation.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 11
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds IN LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14602312
Report Number1818910-2022-10148
Device Sequence Number5064955
Product Code LZO
UDI-Device Identifier10603295168881
UDI-Public10603295168881
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K190344
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L93711
Device Catalogue Number3L93711
Device Lot Number1731630
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer08/24/2022
09/30/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date08/25/2022
10/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OPTION CUP HA 54MM; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-