Model Number 1365-33-000
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient had left hip replaced in 2002.There was no op or post-op report of this case available.Surgeon followed up on the patient's hip replacement because of a small mass that was evident during an ultrasound adjacent to the medial wall of the acetabulum.It was deceived that the ultamet metal liner and metal head articulation could be generating metal debris that could be causing lysis around the acetabular cup.The metal liner, the femoral hip ball was explanted and replaced with a polyethylene liner and a new hip ball.The srom femoral stem, srom sleeve, and pinnacle acetabular cup were left in situ.The lot and catalogue numbers for the implants left in situ and explanted did not have legible numbers to record.There was no surgical delay.Doi: 2002.Dor: (b)(6) 2022.Affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S-ROM M HEAD 36MM +6
Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14602591
Report Number1818910-2022-10155
Device Sequence Number6177472
Product Code LPH
UDI-Device Identifier10603295033509
UDI-Public10603295033509
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K120599
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1365-33-000
Device Catalogue Number136533000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINNACLE MTL INS NEUT36IDX54OD; UNK HIP ACETABULAR CUP PINNACLE; UNK HIP FEMORAL SLEEVE; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
Patient Weight74 KG
-
-