Model Number M0035476300
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device is unavailable to manufacturer.
 
Event or Problem Description
It was reported that the subject coil detached during the procedure when an attempt was made to retract it into the microcatheter to reposition the subject coil.No clinical consequences reported to the patient.
 
Event or Problem Description
It was reported that the subject coil detached during the procedure when an attempt was made to retract it into the microcatheter to reposition the subject coil.No clinical consequences reported to the patient.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject coil device was returned together with the sl-10 microcatheter and un-known guidewire.The sl-10 microcatheter and guidewire were found to be intact.The subject main coil was found to be stretched and detached from the delivery wire and stuck inside the microcatheter due to the dry blood.The coil introducer sheath and the sdw (stent delivery wire) were not returned.An attempt was taken to remove the subject coil from the microcatheter for further analysis.The subject coil device could not be flushed due to dry blood within the microcatheter.The subject coil could not be pushed using returned guidewire.The subject coil was pulled out from the distal end of the microcatheter.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.During analysis, the main coil was found to be stretched and detached from the delivery wire and stuck inside the microcatheter due to dry blood.The delivery wire was not returned.An assignable cause of procedural factors will be assigned to the as reported/as analyzed 'main coil prematurely detached/separated during use' and main coil stretched', as these defects appear to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 6MM X 30CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14602785
Report Number3008881809-2022-00273
Device Sequence Number12290408
Product Code HCG
UDI-Device Identifier04546540676559
UDI-Public04546540676559
Combination Product (Y/N)N
Initial Reporter CountryHU
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2025
Device Model NumberM0035476300
Device Catalogue NumberM0035476300
Device Lot Number23504005
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
-
-