• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 STANDARD 14MM X 50CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 STANDARD 14MM X 50CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036101450
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the subject coil detached during the procedure when an attempt was made to retract it into the microcatheter to reposition the subject coil.No clinical consequences reported to the patient.
 
Event or Problem Description
It was reported that the subject coil detached during the procedure when an attempt was made to retract it into the microcatheter to reposition the subject coil.No clinical consequences reported to the patient.
 
Additional Manufacturer Narrative
The automated manufacturing execution system (mes) has controls in the manufacturing process to ensure the product met specifications upon release.The device was returned together with un-known device.The main coil was found to be kinked/bent and detached/separated.Additionally, the core delivery wire was found to be kinked/ bent.The functional testing could not be performed due to the condition of the returned device.The reported main coil prematurely detached/separated during use was confirmed during the analysis.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The returned device was analyzed.The main coil was found to be prematurely detached/separated and kinked/bent.The coil delivery wire was found to be kinked/bent.An assignable cause of procedural factors will be assigned to the reported and as analyzed ¿main coil prematurely detached/separated during use and to ¿main coil kinked/bent¿, as the issues are associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.A probable cause of handling damage was assigned to the ¿coil delivery wire kinked/bent¿ as the defect appears to be associated with handling of the product or portion of the product during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 STANDARD 14MM X 50CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14603170
Report Number3008881809-2022-00274
Device Sequence Number6146738
Product Code HCG
UDI-Device Identifier04546540711298
UDI-Public04546540711298
Combination Product (Y/N)N
Initial Reporter CountryHU
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2024
Device Model NumberM0036101450
Device Catalogue NumberM0036101450
Device Lot Number23283715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexUnknown
-
-