H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus functional testing could not be performed.There was no way to determine if the device contributed to the reported event.Visual inspection of the customer provided picture shows a foot pedal from a top down view with rust seen around the chassis but provides no complaint related information.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found similar reported events.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|