ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, trailing haptic issue was found.Th optic did not slide through 2.4mm wound.The incision was enlarged and new lens used during initial procedure.No injury to patient was reported.Additional information received and stated that the patient was not hospitalized for the event and there was no patient harm.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A qualified cartridge was indicated.It is unknown if a qualified handpiece and viscoelastic were used.The product investigation could not identify a root cause for the reported complaint.The initial report stated trailing haptic issue.Further information stated the lens would not go thorough the 2.4 incision.It is unknown if a qualified handpiece and viscoelastic were used.The ifu (instructions for use) instructs: the lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Failure to follow this step may cause the lens to advance incorrectly causing delivery issues and/or damage.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14604132
Report Number1119421-2022-01198
Device Sequence Number20375743
Product Code HQL
UDI-Device Identifier00380655095744
UDI-Public00380655095744
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.270
Device Lot Number12664238
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer07/30/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexFemale
-
-