• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON CARIBE LTD. BD BACTEC¿ STANDARD/10 AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON CARIBE LTD. BD BACTEC¿ STANDARD/10 AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING Back to Search Results
Model Number 442027
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: a device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while testing with bd bactec¿ standard/10 aerobic/f culture vials (plastic) a false positive result was obtained.There was no indication that results were reported out, and there was no report of patient impact.The following information was provided by the initial reporter: customer reports that he obtained a false positive result on an apheresis bag during a bactec fx equipment verification procedure.
 
Event or Problem Description
It was reported while testing with bd bactec¿ standard/10 aerobic/f culture vials (plastic) a false positive result was obtained.There was no indication that results were reported out, and there was no report of patient impact.The following information was provided by the initial reporter: customer reports that he obtained a false positive result on an apheresis bag during a bactec fx equipment verification procedure.
 
Additional Manufacturer Narrative
H.6 investigation summary: catalog 442027 batch no.1245169 customer reported a false positive defect.Neither photos nor returned good samples were received.Bd was unable to reproduce customer experience with the bactec product.A false positive response was not observed when retention samples were tested.Batch history records were reviewed, and all testing were within specification for product release.A complaint history review was conducted and only the current complaint was found relating to the incident lot number and the ¿as reported¿ defect code.Users are cautioned in the package insert under limitation of the procedure: ¿a gram-stained smear from culture medium may contain small number of non-viable organisms derived from media constituents, staining reagents, immersion oil, glass slide and specimens used for inoculation.There are many factors that can influence the false positive rate, including blood volume, blood cell counts, environmental factors, and media lot to lot variations.Complaint is unconfirmed based on retention samples and batch history record review.No correctives actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process.Plot/log file review: there is nothing in the log files that is suggesting any voltage or rack issue was present.The unit is functioning as intended.However, the log files showed the voltage outputs ( the readable ones to the computer, not the actual voltages going to the system) were reading significantly off.This may be contributed to a grounding issue, or some other issue with the edb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ STANDARD/10 AEROBIC/F CULTURE VIALS (PLASTIC)
Common Device Name
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14604635
Report Number2647876-2022-00153
Device Sequence Number12053191
Product Code MDB
UDI-Device Identifier00382904420277
UDI-Public00382904420277
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K161810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2022
Device Model Number442027
Device Catalogue Number442027
Device Lot Number1245169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-