| Model Number |
441386 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while testing with bd bactec¿ fx, instrument bottom, packaged false positive results were obtained.This occurred 10 times.Confirmatory testing was performed using gram staining.There was no indication that results were reported out, and there was no report of patient impact.The following information was provided by the initial reporter: the specimens in question were all originally processed in other drawers of the instrument with the specimens being loaded into drawer c and having their protocol length changed to ten days.No instrument issue found.Issue resolved.S/n: (b)(4).Customer problem: (b)(6) called to report having multiple positives from drawer c of this fx instrument.(b)(6) indicated they had three specimens that flagged as positive at or near the same time that all showed no growth when confirmed with gram stain.(b)(6) indicated that they were different bottle types."vials affected (for blood culture only)?" y, ten.Drawer c multiple positives.
|
| |
|
Event or Problem Description
|
|
It was reported while testing with bd bactec¿ fx, instrument bottom, packaged false positive results were obtained.This occurred 10 times.Confirmatory testing was performed using gram staining.There was no indication that results were reported out, and there was no report of patient impact.The following information was provided by the initial reporter: the specimens in question were all originally processed in other drawers of the instrument with the specimens being loaded into drawer c and having their protocol length changed to ten days.No instrument issue found.Issue resolved.S/n: (b)(6) customer problem: s called to report having multiple positives from drawer c of this fx instrument.S indicated they had three specimens that flagged as positive at or near the same time that all showed no growth when confirmed with gram stain.S indicated that they were different bottle types.Vials affected (for blood culture only)? y, ten.Drawer c multiple positives.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: a complaint of "drawer c multiple positives" was received against instrument bottom bactec fx, material no: 441386, serial no: (b)(6).The complaint is not confirmed because no instrument issues were identified and the root cause is "unknown".Review of device history record for instrument serial number, (b)(6) is not required because this complaint does not allege an early life failure or failure at installation and has changed configuration since release from manufacturing due to service repairs/pms.Device was installed on (b)(6) 2013.Samples were not received by quality for investigation and thus, returned material investigation could not occur.Service history review revealed no previous complaints for this issue.Complaints for ¿results¿ issues which included " false positives " complaints are under statistical control for the month of may 2022.The upper control limit was not breached.Quality will continue to monitor the " false positives ¿complaints.No trends were identified.Per baltrmbactecinstraph revision 15, the error related to this complaint is considered user inconvenience, which is an s1; reference row id 1.2c.No new risks or hazards were identified.
|
| |
|