| Model Number |
24674 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 95% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.A 5.0 x 40, 135cm mustang balloon catheter was advanced for dilatation.However, during third inflation at 20 atmospheres for 3 seconds, the balloon ruptured.The device was removed and a pinhole in the center of the balloon was noticed.The procedure was completed with a different device.No patient complications nor injuries were reported and the patient status was good.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 device codes (1): added entrapment of device.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 95% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.A 5.0 x 40, 135cm mustang balloon catheter was advanced for dilatation.However, during third inflation at 20 atmospheres for 3 seconds, the balloon ruptured.The device was removed and a pinhole in the center of the balloon was noticed.The procedure was completed with a different device.No patient complications nor injuries were reported and the patient status was good.It was further reported that the balloon had to be removed with the entire guidewire.
|
| |
|