| Model Number |
FACTOR VIII CHROMOGENIC ASSAY |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An outside of the united states (ous) customer contacted a siemens customer care center (ccc).All reaction kinetics were evaluated correctly by the system software.The issue was limited to results for one patient.A sample specific issue related to the patient's treatment cannot be ruled out as a potential cause of the event.The cause of the event is unknown.The reagent is performing according to specifications.No further evaluation of this device is required.
|
| |
|
Event or Problem Description
|
|
A discordant, falsely elevated factor viii result was obtained on a patient sample on a sysmex cs-5100 system using factor viii chromogenic assay.The discordant result was not reported to the physician(s).A second sample from the same patient that was drawn at the same time was measured for factor viii using a non-siemens assay, recovering lower.It is unknown if this lower result was considered to be correct or if it was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to discordant factor viii result.
|
| |
|