| Model Number |
AED PLUS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would not power up.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical canada.The customer's report was observed during testing.However, the reported problem could not be duplicated.The main board and battery board were replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is correcting information submitted on the previous medwatch reports.Please reference section d4 (serial number) and h4.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical canada; the customer's report was observed and the main board was replaced.The device was recertified and returned to the customer.The main board was returned to zoll medical united states; the customer's report was duplicated and attributed to a shorted transient voltage suppressor diode on the main board.Analysis for reports of this type has not identified an increase in trend.
|
| |
|