• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE REPOSITIONABLE CLIP; SINGLE USE REPOSITIONAL CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE REPOSITIONABLE CLIP; SINGLE USE REPOSITIONAL CLIP Back to Search Results
Model Number HX-202UR
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Perforation (2001)
Date of Event 03/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device referenced in this report has not been returned to olympus for evaluation.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Event or Problem Description
It is reported in the literature article titled ¿prophylactic clip closure for mucosal defects is associated with reduced adverse events after colorectal endoscopic submucosal dissection: a propensity-score matching analysis¿ patients experienced perforation and/or bleeding events during or after procedures using olympus devices.Case with patient identifier (b)(6) reports kd-655q.Case with patient identifier (b)(6) reports hx-610-090l.Case with patient identifier (b)(6) reports hx-202ur.Case with patient identifier (b)(6) reports pcf-h290z1.Case with patient identifier (b)(6) reports pcf-q260ai.Study background and aim it is unclear whether prophylactic endoscopic closure after colorectal endoscopic submucosal dissection (esd) reduces the risk of postoperative adverse events due to variability in lesion characteristics.Therefore, we conducted a retrospective study using propensity score matching to evaluate the efficacy of prophylactic clip closure in preventing postoperative adverse events after colorectal esd methods: this single-center retrospective cohort study included 219 colorectal neoplasms which were removed by esd.The patients were allocated into the closure and non-closure groups, which were compared before and after propensity-score matching.Post-esd adverse events including major and minor bleeding and delayed perforation were compared between the two groups.Results: in this present study, 97 and 122 lesions were allocated to the closure and non-closure groups, respectively, and propensity score matching created 61 matched pairs.The rate of adverse events was significantly lower in the closure group than in the non-closure group (8% vs.28%, p=0.008).Delayed perforation occurred in two patients in the non-closure group, whereas no patient in the closure group developed delayed perforation.In contrast, there were no significant differences in other postoperative events including the rate of abdominal pain; fever, white blood cell count, and c-reactive protein; and appetite loss between the two groups.Conclusions: propensity score matching analysis demonstrated that prophylactic closure was associated with a significantly reduced rate of adverse events after colorectal esd.When technically feasible, mucosal defect closure after colorectal esd may result in a favorable postoperative course.The total number of adverse events after colorectal esd, including delayed perforation and post-esd bleeding, was significantly higher in the non-closure group than in the closure group (28% vs.6%, p < 0.001).Similarly, the proportion of patients with fever was significantly higher in the non-closure group than in the closure group (12% vs.2%, p=0.019).However, there were no significant differences in the proportion of patients with abdominal pain, wbc count, c-reactive protein levels, and appetite loss between the two groups.There is no report of any olympus device malfunction reported in any procedure described in this literature.
 
Event or Problem Description
New information reported by the author/physician: there was no malfunction of any olympus devices during the procedures described in this literature.I do not believe that an olympus device or product specifications are related to the adverse events that occurred.
 
Additional Manufacturer Narrative
This report is being submitted to report the investigation findings and additional information reported by the author/physician.New information is reproted in b5, h6, and h10.The devices referenced in this report were not returned to olympus for physical evaluation.The device history record (dhr) for the complaint devices could not be reviewed since the serial numbers were not provided.Olympus does not ship any device that does not meet all design and safety specifications.It is not possible to conclusively identify the root cause of the reported event.Based on the investigation finding and additional information provided by the author/physician it is likely the event is a complication associated with the surgical procedure.The event occurred after the doctor considered the clinical risk of the surgical procedure.In addition, the author/physician confirmed there was no olympus device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE REPOSITIONABLE CLIP
Common Device Name
SINGLE USE REPOSITIONAL CLIP
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14605021
Report Number8010047-2022-09505
Device Sequence Number20125001
Product Code PKL
UDI-Device Identifier04953170345180
UDI-Public04953170345180
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123601
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Literature,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHX-202UR
Device Lot NumberUNKNOWN(LITERATURE)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-