| Model Number |
DS2C019 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2022-11824.Related manufacturer reference number: 2017865-2022-11827.It was reported that a patient presented in-clinic for an explant procedure.Examination of the patient's lead revealed high capture thresholds.During the procedure and during an attempt to place a left ventricular lead was found to be difficult to remove from the guidewire, and the left ventricular lead was found to be damaged.Additionally, a steerable catheter used was found to have perforated the patient blood vessels, resulting in additional surgical intervention to resolve the complication.The exiting lead was capped on (b)(6) 2022, and implant of the left ventricular lead was abandoned.The patient was stable throughout.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of ¿due to repeated operation, the coronal sinus orifice was interbedded¿.As received only the dilator was returned for analysis.The catheter was not returned.Visual examination did not find any anomalies.
|
| |
|