MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CPS DIRECT UNIVERSAL SLITTABLE OUTER CATHETER; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CPS DIRECT UNIVERSAL SLITTABLE OUTER CATHETER; CATHETER, PERCUTANEOUS Back to Search Results
Model Number DS2C019
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 05/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 2017865-2022-11824.Related manufacturer reference number: 2017865-2022-11827.It was reported that a patient presented in-clinic for an explant procedure.Examination of the patient's lead revealed high capture thresholds.During the procedure and during an attempt to place a left ventricular lead was found to be difficult to remove from the guidewire, and the left ventricular lead was found to be damaged.Additionally, a steerable catheter used was found to have perforated the patient blood vessels, resulting in additional surgical intervention to resolve the complication.The exiting lead was capped on (b)(6) 2022, and implant of the left ventricular lead was abandoned.The patient was stable throughout.
 
Additional Manufacturer Narrative
The reported event of ¿due to repeated operation, the coronal sinus orifice was interbedded¿.As received only the dilator was returned for analysis.The catheter was not returned.Visual examination did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CPS DIRECT UNIVERSAL SLITTABLE OUTER CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14605058
Report Number2017865-2022-11825
Device Sequence Number5065057
Product Code DQY
UDI-Device Identifier05415067004185
UDI-Public05415067004185
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K130257
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberDS2C019
Device Catalogue NumberDS2C019
Device Lot Number7673495
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date09/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-