DEPUY MITEK LLC US CURVED OSTEOTOME; ORTHOPAEDIC OSTEOTOME Back to Search Results
Model Number 285270
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).The lot number was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name and device manufacture date were unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that during an unknown surgery on (b)(6) 2022, it was observed that the curved osteotome device was broken.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in his report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate investigation summary:the complaint device is not being returned, therefore unavailable for a physical evaluation, however 3 photos were provided by the customer.Upon visual inspection of the photos, the osteotome is slightly worn and used, but in expected condition.When inspecting at the handle¿s area, some malleting marks could be observed, the handle¿s cap is broken due to the hard use.Given that lot number was not provided, the manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual evaluation result, this complaint can be confirmed.The possible root cause can be attributed to the constant and rough use.Since this is a reusable device, the constant manipulation between surgeries and repeated sterilization process can lead to a broken handle's cap; however, this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: h4: the device manufacture date was reported as unknown on the initial report; and has been updated accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not availabe for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection, the device showed wear marks, but in expected condition.When inspecting at the handle¿s area, some malleting marks could be observed, the handle¿s cap is broken due to the hard use.The welding near the connection point was broken.A manufacturing record evaluation was performed for the finished device lot number:18d01, and no nonconformances were identified.According with the visual evaluation result, this complaint can be confirmed.The possible root cause can be attributed to the constant and rough use.Since this is a reusable device, the constant manipulation between surgeries and repeated sterilization process can lead to a broken handle's cap; the broken condition can related to a welding issue; when applying excessive force after multiples uses affecting the connection point.However, this cannot be conclusively determined.As per ifu, inspect the instruments before use for integrity and compatibility in order to ensure proper functionality and safety.Prior to sterilization, instruments should be cleaned by either manual or automated means.Care should be taken to avoid the two osteotomes hitting one another as damage to the cutting edges can occur.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURVED OSTEOTOME
Common Device Name
ORTHOPAEDIC OSTEOTOME
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14605161
Report Number1221934-2022-01645
Device Sequence Number5164825
Product Code LXH
UDI-Device Identifier10886705018000
UDI-Public10886705018000
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number285270
Device Catalogue Number285270
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer06/14/2022
06/22/2022
06/29/2022
07/01/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date06/15/2022
06/24/2022
06/30/2022
07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1