MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*SKIN STAPLER-WIDE; REMOVABLE (SKIN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*SKIN STAPLER-WIDE; REMOVABLE (SKIN) Back to Search Results
Model Number PXW35
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure, the staples took off.The staples did not hold.Passing the hand on them made them fall.Surgical delay.A suture by thread had been performed instead.No patient consequence.
 
Additional Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch/lot number, and no non-conformances were identified.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 07/22/2022.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the pxw35 device was returned with no apparent damage, fully functional.In addition, an opened packaging was received.When tested for functionality the device fired and formed the remaining 25 staples as intended.The device fired without any difficulties and the staples were noted to have the proper box shape.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: with two tissue forceps, pick up each wound edge individually and approximate the edges.Or, with one tissue forcep, pull skin edges together until edges evert.Or, apply tension to either end of the incision, such that the tissue edges begin to approximate themselves.One forcep can be used to ensure that the edges are everted.Position the instrument over the everted skin edges, aligning the instrument arrow with the center of the incision at a 50° - 60° angle from the skin.Squeeze the trigger until you touch plastic trigger to plastic handle (plastic to plastic), then release the trigger and move the instrument off the incision.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
REMOVABLE (SKIN)
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
4188 lexington avenue north
shoreview 55126
MDR Report Key14605452
Report Number3005075853-2022-03645
Device Sequence Number14980854
Product Code GDT
UDI-Device Identifier10705036002673
UDI-Public10705036002673
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPXW35
Device Catalogue NumberPXW35
Device Lot Number666A09
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-