Model Number SV-494
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.A review of the dhr for the lot was performed and no issues were noted during manufacturing.Retain samples from the same lot were tested for finished goods testing, including tensile strength and needle attachment tesing.Test results show that all retains passed.Investigations performed did not reveal any issues with the product and the report could not be substantiated.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical, or its employees, that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr part 803.If additional information is provdided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
 
Event or Problem Description
According to the reporter, "during an open procedure, the needle of the suture detached from the thread and fell into patient's cavity.The surgery was extended by 30 minutes or more to find the fallen needle but then it was retrieved and procedure was completed.No patient injury.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELOSORB FAST
Common Device Name
BRAIDED FAST ABSORBABLE SUTURE
Manufacturer (Section D)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer (Section G)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer Contact
edwin anderson
825 ne 25th ave
portland, OR 97232
5035178001
MDR Report Key14605650
Report Number3006981798-2022-00011
Device Sequence Number6146801
Product Code GAM
UDI-Device Identifier10884521151130
UDI-Public10884521151130
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K100461
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberSV-494
Device Lot Number21033003FX
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Initial Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-