• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0102
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure in the anterior communicating artery (acom) using penumbra smart coils, a smart coil detachment handle (handle), and a non-penumbra microcatheter.During the procedure, the physician advanced a smart coil into the target vessel using the microcatheter and attempted to detach it using the handle.However; the smart coil failed to detach.After many attempts to detach the smart coil using the handle, the physician then manually detached the smart coil into the target vessel.It was reported that the same issue occurred with two additional smart coils and the two smart coils were also successfully manually detached in the target vessel.Next, the physician advanced a fourth smart coil into the target vessel and attempted to detach it using the handle.However; the smart coil failed to detach.After many attempts to detach the smart coil using the handle and manual detachment, part of the smart coil was pulled outside of the aneurysm and into the parent vessel.While attempting to retract the smart coil into the microcatheter, the smart coil unintentionally detached with part of the smart coil in the parent vessel.Therefore; the physician used a new non-penumbra microcatheter and a non-penumbra stent retriever device to retrieve the detached smart coil.It was noted that the smart coil was elongated and ¿stretched¿ upon removal.The procedure ended at this point.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14606249
Report Number3005168196-2022-00267
Device Sequence Number20083835
Product Code HCG
UDI-Device Identifier00814548015613
UDI-Public00814548015613
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/06/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2024
Device Model Number400SMTHXSFT0102
Device Catalogue Number400SMTHXSFT0102
Device Lot NumberF93755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Initial Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-