Model Number PLC121000
Medical Device Problem Code Inadequate User Interface (2958)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 10/20/2021
Type of Reportable Event Death
Event or Problem Description
On (b)(6) 2021, this patient underwent endovascular treatment of an abdominal aortic aneurysm and was implanted with gore® excluder® conformable aaa endoprostheses (excc).After gaining access via bilateral femoral cutdowns, sheaths were advanced into the mid and proximal infrarenal aorta.The excc was advanced thru a 16 french sheath and placed at the level of the renal arteries.Angiography was performed and the device was deployed without incident, however the contralateral gate was noted to be compressed.Cannulation of the gate was performed without difficulty and a pigtail was advanced and spun to confirm that the wire was in the body of the graft.A contralateral limb (plc121000/(b)(4)) was then advanced after the sheath and dilator were placed into the gate of the trunk.The sheath was drawn back below the level of the left hypogastric artery and deployed.The trunk was then fully deployed and an extension limb (plc121200/(b)(4)), which was marked after doing an injection of contrast thru the sheath was place to the level of the right hypogastric artery.The proximal part of the trunk was ballooned with a coda balloon and kissing 12 mm pta balloons were used in the limbs.Then bilateral 10mm diameter luminexx¿ self-expanding stents were implanted in the external iliac arteries.At this time the patient¿s blood pressure was noted by anesthesia to be dropping.Angiography was performed and no leak was seen however a dissection of the left iliac artery (lia) distal to the graft was determined.A 13x10 gore® viabahn® endoprosthesis with heparin bioactive surface was implanted in the lia with intentional coverage of the hypogastric artery.Post placement of viabahn® the patient was transfused and the blood pressure improved.Approximately 5 minutes later the patient¿s blood pressure dropped again and the physician chose to open the patient¿s belly, the coda balloon was advanced to the level of the visceral vessels and inflated, which stabilized the blood pressure.Once exposure was done, a rupture of the aorta at the level of the limbs approximately 15-17 mm distally was determined.The patient was converted to open and an aortobifemoral bypass was performed.The patient tolerated the procedure in stable condition.It was reported that the physician noted that patient was not a good surgical candidate.Concerns noted post review of the films were the patient¿s short proximal aortic neck (10mm -12 mm in length) a small distal aorta (15mm -17 mm in diameter over a 3 cm length) and calcification of the access vessel coupled with the diameter of the access vessel bilaterally (5.5mm -6 mm diameter) bilaterally.After all concerns were discussed with the patient the physician chose to proceed with evar.The dissection was thought to have been due to the calcification throughout the patient¿s access vessels and the rupture of the distal aorta is believed to have been caused by angioplasty with the kissing balloons.On (b)(6) 2021, it was reported that the patient expired.The physician stated the cause of death was multisystem organ failure.The physician stated the multisystem organ failure could have been caused by the emergent conversion due to the rupture of the aorta when ballooning the two limbs down close to the aortic bifurcation.She ended up having issues due to prolonged anesthesia of almost 8 1/2 to 9 hours.As well the patient was not an ideal surgical candidate in the first place and that was the reason for attempting the endovascular repair.
 
Additional Manufacturer Narrative
Images were provided to gore for evaluation and showed the following: images provided for evaluation are pre-implant cta.Diameter just distal to the left renal measures ~ 19 mm.Diameter ~ 5 mm distal to the left renal measures ~ 17 mm.Diameter ~ 10 mm distal to the left renal measures ~ 16 mm.Diameter ~ 15 mm distal to the left renal measures ~ 30 mm.There appears to be a reverse taper present at the proximal neck.There does not appear to be 15 mm of aortic neck available.Diameters in the 3.3 cm proximal to the aortic bifurcation range from 9.6 mm ¿ 16.6 mm.Distal waist diameters measure < 18 mm.Diameters in the rci range from 5.4 mm ¿ 7.2 mm (just proximal to the riia).Diameters in the rei range from 5.1 mm ¿ 6.4 mm.Diameters in the lci range from 7.1 mm ¿ 8.3 mm.Diameters in the lei range from 5.3 mm ¿ 6.4 mm.According to the instructions for use (ifu) for the gore® excluder® conformable aaa endoprosthesis is intended to exclude the aneurysm from the blood circulation in patients diagnosed with infrarenal abdominal aortic aneurysm (aaa) disease and who have appropriate anatomy as described below: infrarenal aortic neck treatment diameter range of 16 ¿ 32 mm and a minimum aortic neck length of 15 mm.Additionally, the ifu states, adverse events that may occur and/or require intervention include, but are not limited to: vascular spasm or vascular trauma (e.G., aorta dissection, aorta damage, ilio-femoral vessel dissection, bleeding, rupture, death).Patient medical history includes but is not limited to: cad.Patient medications include but are not limited to: unspecified medications for treatment of hypertension and cad.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14607624
Report Number3013164176-2022-01398
Device Sequence Number4456426
Product Code MIH
UDI-Device Identifier00733132618613
UDI-Public00733132618613
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/25/2024
Device Model NumberPLC121000
Device Catalogue NumberPLC121000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2021
Initial Report FDA Received Date06/06/2022
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Life Threatening; Other; Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexFemale
Patient Weight74 KG
-
-