• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII SPC PRI SPACER BLOCK; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII SPC PRI SPACER BLOCK; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71441265
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a tka surgery, in two (2) gii spc pri spacer block the plastic locking mechanism was fractured/cracked after many uses.The procedure was completed, without delay, using the same device.Patient was not harmed.
 
Additional Manufacturer Narrative
H3, h6: the devices were not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.The devices are reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable, damage may occur during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the devices to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual products involved, our investigation could not proceed.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII SPC PRI SPACER BLOCK
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14609036
Report Number1020279-2022-02758
Device Sequence Number5367156
Product Code HRY
UDI-Device Identifier00885556034354
UDI-Public00885556034354
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71441265
Device Catalogue Number71441265
Device Lot Number08FM09335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date06/06/2022
Supplement Report FDA Received Date06/30/2022
Date Device Manufactured07/12/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-