• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZA9003
Medical Device Problem Codes Break (1069); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that as the intraocular lens (iol) was unfolding in patient's operative eye, just after implantation, it was noted that the trailing haptic was bent causing the lens not to center in the eye.The lens was cut in half and removed.Another lens was implanted without issues.No other information was provided.
 
Additional Manufacturer Narrative
Age or date of birth, weight and ethnicity: unknown/ asked but not available.If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information received from the facility revealed the following information.The doctor is dr.(b)(6), the eye affected was left eye.There was no patient injury and no intervention was required.The procedure was completed using another lens of same model and diopter.Also, the following fields are updated to reflect additional information.Section a2: age/date of birth: (b)(6), 1954.Section a3: gender: female.Section a4: patient weight: 180 pounds.Section a5: ethnicity/race: hispanic/ latino, white.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: returned to manufacturer on: sep 2, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint lens was received cut in half and with a detached haptic.The remaining haptic was observed to be bent.The lens was cleaned and, no further issues were observed.Conclusion: the complaint issue of "haptic damaged" was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14613773
Report Number3012236936-2022-01452
Device Sequence Number4509145
Product Code HQL
UDI-Device Identifier05050474528963
UDI-Public(01)05050474528963(17)270208
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZA9003
Device Catalogue NumberZA90030175
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer06/22/2022
09/25/2022
Initial Report FDA Received Date06/07/2022
Supplement Report FDA Received Date07/12/2022
10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight82 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-