Model Number 2090
Medical Device Problem Codes No Display/Image (1183); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment of the broken power cord bay however it was noted that there was an intermittent failure with the stylus and cursor and the overlay to xy printed circuit board (pcb) connection was reseated and the stylus calibrated.It was noted that the upper case was broken at the carrying handle.Reconfigured hard drive, reloaded and updated software and the device then passed all final functional and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the cord cover lid required repair.It was noted that the programmer required a software update.It was noted that the programmer subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14617096
Report Number2182208-2022-01848
Device Sequence Number14566873
Product Code KRG
UDI-Device Identifier00613994434937
UDI-Public00613994434937
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2022
Initial Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2067 RADIO FREQUENCY HEAD
Patient SexUnknown
-
-