Model Number 16-02-85
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t had no flow on patient circuit 1 and was not warming properly.The issue was identified during maintenance.There was no patient involvement.
 
Manufacturer Narrative
Through follow-up communication livanova learned that the reported issue affected cardioplegia circuit only and not patient circuit.Pump malfunction and heating malfunction on cardioplegia circuit are not likely to result is death or serious injury since cardioplegia solution is delivered in small volumes and at regular intervals allowing the user to perfom heating through alternative methods or to use another circuit of the device.Thus, temperature management on cardioplegia side is not critical in clinical practice and the reported event is assessed as not reportable.In addition, livanova identified that this complaint was already captured in livanova complaint database.Therefore, this specific case has been voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14618486
MDR Text Key293457636
Report Number9611109-2022-00291
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/11/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/23/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-