| Model Number |
T401161 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
|
| |
|
Event or Problem Description
|
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the locking mechanism was broken.It is still unknown whether there was a delay in the procedure, whether the surgery was completed successfully and whether there was blood loss.Due to the unknown information, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.The product was changed out.
|
| |
|
Event or Problem Description
|
|
Additional information received from sales associate, that there was no delay in the procedure.The surgery was completed successfully and there was no blood loss as a result of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on jun 7, 2022.Upon further investigation of the reported event, the following information is new and/or changed: b5 (added describe event or problem); d5 (operator of device); e1 (initial reporter - email address); e2 (health professional); g2 (report source - health professional); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information) ; h6 (identification of codes 2645, 4582, 11).H6: health effect - impact code: 2645 - no patient involvement.Health effect - clinical code: 4582 - no clinical signs, symptoms or conditions.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d9 (device availability - added date returned to manufacturer).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 (device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 3331, 3259, 4307) type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 3331 - analysis of production records investigation finding: 3259 - improper physical structure investigation conclusions: 4307 - cause traced to component failure the affected sample was inspected upon receipt to confirm a broken lock plate.Mild corrosion was observed on the device, including the lock plates.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|