| Model Number |
3 LEADS SET DISPOSABLE, BEDSIDE IEC RD |
| Medical Device Problem Codes |
Overheating of Device (1437); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported to philips that the electrode l of the ecg cable began to burn the patient.The device was in use at time of event, a minor burn was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to philips.Visual inspection of the lead set found that the lead set has burn damaged on wire for the yellow grabber.It cannot be confirmed if the cable is fatigued, or damage was caused by tension on the lead set.The lead set could not be functionally due to hazard of exposed wire.The customer has ordered the replacement part to resolve the issue.The device will be held in the mcs lab for further review if required.
|
| |
|
Event or Problem Description
|
|
It was reported to philips that the electrode l of the ecg cable began to burn the patient.The device was in use at time of event, a minor burn was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was updated as part of a retrospective review based on the available information.
|
| |
|
Event or Problem Description
|
|
It was reported while monitoring ecg, the patient developed a burn from the ecg cable.The burn was described as minor, for which unknown medical attention was provided.Images of the burn damaged ecg cable and the burn on the patient's skin were received.No additional information was provided.
|
| |
|