• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG RIVACOR 5 HF-T QP DF4 IS4 PROMRI; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG RIVACOR 5 HF-T QP DF4 IS4 PROMRI; ICD Back to Search Results
Model Number 429561
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock (2072)
Date of Event 10/08/2021
Type of Reportable Event Serious Injury
Event or Problem Description
After an implantation period of approx.12 months, it was observed that the patient received an inappropriate shock from the crt.Device remains implanted.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The icd was not returned for analysis.The analysis is therefore based on the inspection of the quality documents associated with the manufacture of this particular device as well as on the returned device data.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.The final acceptance test proved the icd functions to be as specified.The returned device data have been analyzed.The analysis did not show any anomalies.A vf and an svt episode were detected on october 8th, 2021 at 17:50 h and 17:51 h, respectively.The vf detection was appropriate and resulted in a 40 j shock delivery.Right after the shock delivery an af occurred and subsequently the svt episode without shock delivery.The icd functioned appropriately, both regarding detection and therapy delivery.In conclusion, the device was not returned for analysis.The review of the quality documents confirmed a regular device manufacturing.The analysis of the returned device data revealed no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIVACOR 5 HF-T QP DF4 IS4 PROMRI
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key14620995
Report Number1028232-2022-02880
Device Sequence Number20112682
Product Code NIK
UDI-Device Identifier04035479156879
UDI-Public04035479156879
Combination Product (Y/N)N
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number429561
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date06/07/2022
Supplement Report FDA Received Date07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-