• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUWAVE MEDICAL, INC. NEUWAVE PR XT PROBE 20CM 15GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUWAVE MEDICAL, INC. NEUWAVE PR XT PROBE 20CM 15GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES Back to Search Results
Model Number PR20XT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Additional information was requested and the following was obtained: did the patient experience any adverse consequences due to this issue? if yes, please specify and inform what was done to help the patient.No, the instrument was unpacked and immediately replaced.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.
 
Event or Problem Description
It was reported that during an unknown ablation there was a broken needle.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 7/14/2022.Investigation summary; since this occurred in switzerland, there is no call home available.Probe nm21nhc45921 (0 uses) was investigated at neuwave.The returned probe's lemo nut was disconnected and missing.This causes an incomplete connection with the pdm, therefore the probe is unable to register with the system.It may look as if no probe is connected.Manufacturing also investigated the issue and concluded this probe was built by an employee that is no longer with neuwave.The probe did not have enough adhesive holding the lemo nut in place.A re-training was completed in manufacturing to go over this issue.Lot ml21086828 was reviewed (in process sn (b)(6)).Manufacture date: 9/13/21.Expiration date: 4/30/25.The cannula for probe sn (b)(6) failed the continuity test.This cannula was replaced.No further problems were seen during the manufacturing and testing of this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUWAVE PR XT PROBE 20CM 15GA
Common Device Name
SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES
Manufacturer (Section D)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
Manufacturer (Section G)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
Manufacturer Contact
orla o'mahony
3529 anderson st
madison, WI 53704
6107428552
MDR Report Key14621229
Report Number3008769756-2022-00029
Device Sequence Number5367462
Product Code NEY
UDI-Device Identifier00853719006425
UDI-Public00853719006425
Combination Product (Y/N)N
PMA/510(K) Number
K160936
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPR20XT
Device Catalogue NumberPR20XT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer06/21/2022
Initial Report FDA Received Date06/07/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-