• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAZOR ROBOTICS LTD MAZOR X SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAZOR ROBOTICS LTD MAZOR X SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number TPL0059
Device Problem Imprecision (1307)
Patient Problems Nerve Damage (1979); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/26/2022
Event Type  Injury  
Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that there was a deviation during a l4/l5 case.The left l4 screw was medial and breached the canal.Navigation was used with the dilator, but not when placing the screws.The surgical system was mounted to the patient using a schanz pin placed in the psis.Accuracy was not checked throughout the procedure.The surgeon did not think the deviation was due to the guidance system and they thought the retractor pushed the cannula medially.The amount of deviation was unknown.The use of the guidance system was aborted and the screw was repositioned using navigation.The procedure was delayed less than an hour.
 
Manufacturer Narrative
H3: a software analysis was initiated to determine the probable cause of the issue.The investigating team concluded the root cause of the deviation of left l4 was due to soft-tissue pressure being applied on the tools while instrumenting, which resulted in medially deviated trajectories.However, as left l4 was the last trajectory executed, a platform or patient shift cannot be ruled out.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAZOR X SYSTEM
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS  3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS   3079567
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14621759
MDR Text Key293470688
Report Number3005075696-2022-00048
Device Sequence Number1
Product Code OLO
UDI-Device Identifier07290109183213
UDI-Public07290109183213
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/26/2022
Initial Date FDA Received06/07/2022
Supplement Dates Manufacturer Received08/08/2022
Supplement Dates FDA Received08/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age71 YR
Patient SexMale
Patient Weight100 KG
-
-