Catalog Number RBYPOD6
Medical Device Problem Codes Material Deformation (2976); Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure in the aorta artery using pod coils, pod packing coils (pod pc), and a lantern delivery microcatheter (lantern).During the procedure, after advancing a pod coil as the first coil approximately several centimeters out the distal end of the lantern, the physician noticed the pod would not advance any further and decided to remove the pod coil.However, while retracting the pod coil, the physician experienced resistance and kinked the proximal end of the pod coil.Consequently, the pod coil unintentionally detached with the majority portion of the detached coil inside the lantern.Therefore, the physician used a hemostat to clamp the lantern containing the detached pod coil and removed both the lantern and coil together as a unit.The physician then inserted a new lantern and placed a new pod coil.Subsequently, the physician inserted a pod pc into the lantern; however, while advancing a pod pc through the lantern, the physician experienced resistance and, consequently, the physician kinked the pusher assembly of the pod pc.Therefore, the pod poc was removed.The procedure was completed using new pod pcs and the same lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14621832
Report Number3005168196-2022-00271
Device Sequence Number14580124
Product Code HCG
UDI-Device Identifier00814548019611
UDI-Public814548019611
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/07/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPOD6
Device Lot NumberF113040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Initial Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age86 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-