Model Number 1010480-H
Medical Device Problem Codes Product Quality Problem (1506); Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat 100% stenosed lesion in the mid right coronary artery (mrca) with moderate calcification and mild tortuosity.The 014 hi-torque (ht) pilot guide wire (gw) was advanced to the target lesion; however a non-abbott balloon dilatation catheter (bdc) was advanced on the gw with difficulty.The bdc became stuck 30mm from the proximal point from the distal edge of the gw.Therefore the gw and bdc were removed together from the patient.A bump was observed between the core and polymer cover of the gw.There was no adverse patient effects and no clinically significant delay in the procedure.The procedure was completed with another ht pilot gw and with the bdc for pre-dilatation followed by stent deployed.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.As there was no reported damage noted to the device during the inspection prior to use, it is possible that inadvertent mishandling and/or interaction with the moderately calcified, mildly torturous and 100% stenosed anatomy/other devices resulted in compromising the guide wire resulting in the reported product quality problem irregular appearance (bump); thus resulting in the reported difficult to advance and difficult to remove; however this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE PILOT GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14621925
Report Number2024168-2022-06049
Device Sequence Number12047113
Product Code DQX
UDI-Device Identifier08717648073854
UDI-Public08717648073854
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number1010480-H
Device Catalogue Number1010480-H
Device Lot Number1052871
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date06/07/2022
Supplement Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-