• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TEPHA, INC. - 3005670760 GALAFLEX; BREAST - ABSORBABLE MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TEPHA, INC. - 3005670760 GALAFLEX; BREAST - ABSORBABLE MESH Back to Search Results
Model Number 100035-15
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 06/12/2015
Type of Reportable Event Serious Injury
Event or Problem Description
Per clinical trial (b)(4): on (b)(6) 2015, patient underwent bilateral augmentation mastopexy using a full vertical and horizontal t incision.1 piece of galaflex was used.Concomitant procedure included excision of right abdominal scar.On (b)(6) 2015, allergic reaction all over body, was reported.Patient was treated with medication (stopped keflex and percocet and placed on medrol dose pack).Patient went to er with lip swelling and hives on (b)(6) 2015.Treated with hydroxyzine and famotidine.All resolved without sequelae on (b)(6) 2015.Review by medical affairs identifies the adverse event as likely related to the device and serious.
 
Additional Manufacturer Narrative
Based on the information provided, no conclusions can be made.As reported, patient was diagnosed with post-op complication including allergic reaction.Medical affairs has assessed the patient¿s postoperative course as being serious and likely related to the study device.Review of the device history records confirmed the device was manufactured to specification.This mdr represents the ae of allergic reaction.An additional mdr was submitted to represent to ae of swelling and pain.Note: this mdr report is related to a retrospective study conducted to proactively gather real-world evidence on the use and safety of the p4hb product family, including galaflex scaffold, galaflex 3d, and galaflex 3dr.All safety events from the study are adjudicated by the bd medical monitor for an independent assessment of severity and relatedness.In an abundance of caution, events for which relatedness cannot be ruled out are processed for mdr reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAFLEX
Common Device Name
BREAST - ABSORBABLE MESH
Manufacturer (Section D)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer (Section G)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14622023
Report Number1213643-2022-00385
Device Sequence Number19947388
Product Code OOD
UDI-Device Identifier00855920006034
UDI-Public(01)00855920006034
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K140533
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2017
Device Model Number100035-15
Device Catalogue NumberGP0408
Device Lot Number140411
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Initial Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/18/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age32 YR
Patient SexFemale
Patient Weight57 KG
-
-