| Model Number |
6900PTFX29 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Gradient Increase (1270); Perivalvular Leak (1457)
|
| Health Effect - Clinical Codes |
Angina (1710); Dyspnea (1816); Insufficient Information (4580)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with a 29mm 6900ptfx pericardial valve in the tricuspid position was placed under consideration for a valve-in-valve procedure after an implant duration of four (4) years, 11 months due to tricuspid regurgitation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Based on the information available, the device was intentionally used outside the way it was designed and intended.The performance, safety, efficacy, longevity, and durability of rings and bioprosthetic valves have not been determined when the product is used outside the scope of its intended use and/or its studied population.Regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: updated: b4, b5, g3, g6, h2, h6.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 29mm 6900ptfx pericardial valve in the tricuspid position was placed under consideration for a valve-in-valve procedure after an implant duration of four (4) years, 11 months due to stenosis, moderate-severe central regurgitation with an area of pvl, and mean gradient of 7mmhg.The patient initially presented with shortness of breath, chest pain, and dyspnea.The ttvr procedure was performed after an implant duration of five (5) years, two (2) months resulting in the implant of a 29mm 9600tfx transcatheter valve.Post procedure mean gradient was 1mmhg with no central regurgitation with persistent small area of pvl.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information available, the device was intentionally used outside the way it was designed and intended.The performance, safety, efficacy, longevity, and durability of rings and bioprosthetic valves have not been determined when the product is used outside the scope of its intended use and/or its studied population.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcified degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.Regurgitation is considered to be a perivalvular leak (pvl) if a turbulent eccentric jet originates between the bioprosthetic sewing ring and the annulus.Pvl can occur in the mitral and aortic position for similar reasons.In the early postoperative period, the highest incidence of pvl has been seen in patients developing infective endocarditis, which is most likely attributed to inadequate peri-operative antibiotic prophylaxis or nosocomial infection.Annular calcification is also a risk factor for the development of peri-operative pvl as the bioprosthesis may not seat properly after debridement.Technique related factors, such as incorrect valve sizing, have been shown to contribute to the development of pvl.Anatomical factors may create difficulty seating the bioprosthetic valve resulting in pvl.The anatomy of the annulus may induce mechanical stresses along the rigid bioprosthetic ring which can influence long-term valve performance and durability.A diseased or rigid annulus can potentially increase the mechanical stress on the prosthetic valve, leading to pvl.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
|
| |
|