| Model Number |
02.221.123 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Therapy date (b)(6) 2022.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, 5.0 va locking screw would not thread into the plate in two of the distal interlocking holes.The correct drill guide and drill bit were used to prepare for insertion but screws stripped out in both holes.Another plate was placed with same screws with no issue.There was no reported surgical delay.There were no patient consequences.This report involves (1) 3.5/4.5mm va ppfx prox femur plate/10holes/left.This is report 1 of 3 for (b)(4).
|
| |
|
Search Alerts/Recalls
|