MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an arthroscopy, while the patient was anesthetized, after switching on the light source console and plugging in the cable, it started to produce smoke from the sides and behind.The error message on the display was: ¿lamp fail.E13".The procedure was completed using a competitor device.An unspecified delay was reported.No patient complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H6: the reported device was not returned to the designated complaint facility for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include failure of a concomitant device.A review of device records shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be re-opened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE, 500XL, XENON
Common Device Name
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14622392
Report Number1643264-2022-00194
Device Sequence Number4949368
Product Code FFS
UDI-Device Identifier03596010643070
UDI-Public03596010643070
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K994084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200568
Device Catalogue Number72200568
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date06/07/2022
Supplement Report FDA Received Date09/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-